EST
  • Home
  • Archives
  • Volume 9
  • Number 1
  • Performance of SuperSTAT Method of Non Invasive Blood Pressure (NIBP) Measurement Technique at Different Inflation Levels

Call for paper
April Edition 2017

International Journal of Applied Information Systems solicits high quality original research papers for the
March 15, 2017
April 2017 Edition of the journal.
The last date of research paper submission is
March 15, 2017
SUBMIT YOUR PAPER

Number 1

Performance of SuperSTAT Method of Non Invasive Blood Pressure (NIBP) Measurement Technique at Different Inflation Levels

User Rating: 0 / 5

Star InactiveStar InactiveStar InactiveStar InactiveStar Inactive
 

PrintEmail

journal image
 Download
863
  • Prakash Dirshanapu and Jatoth Ravi Kumar 2015. Performance of SuperSTAT Method of Non Invasive Blood Pressure (NIBP) Measurement Technique at Different Inflation Levels. International Journal of Applied Information Systems. 9, 1 (June 2015), 14-17. DOI=http://dx.doi.org/10.5120/ijais451356
  • @article{10.5120/ijais2017451568,
    author = {Prakash Dirshanapu and Jatoth Ravi Kumar},
    title = {Performance of SuperSTAT Method of Non Invasive Blood Pressure (NIBP) Measurement Technique at Different Inflation Levels},
    journal = {International Journal of Applied Information Systems},
    issue_date = {June 2015},
    volume = {9},
    number = {},
    month = {June},
    year = {2015},
    issn = {},
    pages = {14-17},
    numpages = {},
    url = {/archives/volume9/number1/748-1356},
    doi = { 10.5120/ijais15-451356},
    publisher = { xA9 2015 by IJAIS Journal},
    address = {}
    }
    
  • %1 451356
    %A Prakash Dirshanapu
    %A Jatoth Ravi Kumar
    %T Performance of SuperSTAT Method of Non Invasive Blood Pressure (NIBP) Measurement Technique at Different Inflation Levels
    %J International Journal of Applied Information Systems
    %@ 
    %V 9
    %N 
    %P 14-17
    %D 2015
    %I  xA9 2015 by IJAIS Journal
    

Abstract

The purpose of this paper is to analyze the performance of oscillometric based super STAT method which is used to determine automatic Non Invasive Blood Pressure (NIBP). Mainly four target cuff pressures are chosen in the range of 115mmHg to 160mmHg and at each target cuff pressure the NIBP parameters are measured from 10 numbers of subjects which are performed on left and rights hands and recorded. Eventually the correlation graph is plotted for systolic, diastolic, MAP (Mean Arterial pressure) and pulse readings using statistical analysis tools. The results of the correlation plot states that systolic, diastolic and MAP are moderately correlated at low target inflation cuff pressure and in the range of 135mmHg to 160mmHg cuff inflation pressure the correlation is high between left and right hand NIBP determinations. Pulse measurements are highly correlated at all the cuff inflation levels. Thus the super STAT algorithm gains better accuracy over basic oscillometric method of NIBP determination.

References

  1. Genc Yasemin, Altunkan Sekip, Kiline Oguzhan, Altunkan Erkan. Comparative study on auscultatory and oscillometric methods of ambulatory blood pressure measurements in adult patients Blood Pressure Monitoring 2008; 13: 29-35.
  2. Chio Shiu-Shin, Urbina Elaine M. , orotkoff sound versus oscillometric cuff sphygmomanometers: comparison between auscultatory and DynaPulse blood pressure measurements. Journal of the American Society of Hypertention 2011; 5(1): 12- 20.
  3. Mohamad Forouzanfar, Hilmi R. Dajani, Voicu Z. Groza, Miodrag Bolic, Sreeraman Rajan, and Izmail BatkinRatio Independent Blood Pressure Estimation by Modeling the Oscillometric Waveform Envelope, VOL. 63, NO. 10, OCTOBER 2014.
  4. Kiers H. D. , Hofstra J. M. , Wetzels J. F. M. , Oscillometric blood pressure measurements differences between measured and calculated mean arterial pressure. The Journal of Medicine 2008; 66(11): 474-479.
  5. Bur A, Hirschl MM, Herkner H, Oschatz E, Kofler J, WoisetschlxE4ger C, Laggner AN. Accuracy of oscillometric blood pressure measurement according to the relation between cuff size and upper-arm circumference in critically ill patients, 2000 Feb; 28(2):371-6.

Keywords

Accuracy, Cuff, Diastolic, NIBP, Super STAT Target Inflation, Systolic

Index Terms

Computer Science
Information Sciences